DWG 04 · Work order WO-1191

WO-1191A-308

In repaircritical severityon BME-02011 · ICU ventilator
  1. Reported
  2. Under investigation
  3. Assigned
  4. In repair
  5. Awaiting parts
  6. Resolved
  7. Closed

01Report

High-priority "Exp. valve fault" alarm during ventilation; patient moved to bed 6 ventilator.

Reported
05 Nov 22:10
Reporter
ICU night RT
Error code
A-308
Assigned to
Rafael Alvarez
Device status
Awaiting repair
Location
ICU, bed 2

Symptoms

Expiratory valve fault alarm, PEEP drifting 2-3 cmH2O above set.

02Update ticket

03Audit history

3 entries
  1. 06 Nov 07:20 · Rafael Alvarez

    StatusAssignedIn repair

    Membrane torn at the seat. Replacing from stock.

  2. 05 Nov 22:31 · Rafael Alvarez

    StatusReportedAssigned

    On-call engineer took the ticket.

  3. 05 Nov 22:10 · ICU night RT

    StatusReported

    Reported from ICU bed 2. Patient changed to backup ventilator.