DWG 04 · Work order WO-1187

WO-1187E-21

Awaiting partshigh severityon BME-04217 · Volumetric infusion pump
  1. Reported
  2. Under investigation
  3. Assigned
  4. In repair
  5. Awaiting parts
  6. Resolved
  7. Closed

01Report

Downstream occlusion alarm with the line clear. Pump pulled from bedside.

Reported
03 Nov 07:42
Reporter
4 West charge nurse
Error code
E-21
Assigned to
Rafael Alvarez
Device status
Awaiting repair
Location
4 West, bay 12

Symptoms

Alarm within 2 minutes at 25 mL/h; line patent, no kinks.

alarm-screen.jpg

02Update ticket

03Audit history

5 entries
  1. 03 Nov 11:15 · Rafael Alvarez

    StatusUnder investigationAwaiting parts

    Sensor assembly 81-2210 ordered, ETA 7 Nov. Loaner BME-04230 issued.

  2. 03 Nov 10:52 · Rafael Alvarez

    Investigation note

    Downstream pressure sensor reads 40% high in service mode. Matches KB-112.

  3. 03 Nov 10:30 · Rafael Alvarez

    StatusAssignedUnder investigation

    Alarm reproduces on the test set.

  4. 03 Nov 08:05 · Dana Whitfield

    StatusReportedAssigned

    Assigned to Rafael Alvarez. Two spares left on the unit.

  5. 03 Nov 07:42 · 4 West charge nurse

    StatusReported

    Reported from 4 West, bay 12.