DWG 04 · Work order WO-1187
WO-1187E-21
Awaiting partshigh severityon BME-04217 · Volumetric infusion pump
- Reported
- Under investigation
- Assigned
- In repair
- Awaiting parts
- Resolved
- Closed
01Report
Downstream occlusion alarm with the line clear. Pump pulled from bedside.
- Reported
- 03 Nov 07:42
- Reporter
- 4 West charge nurse
- Error code
- E-21
- Assigned to
- Rafael Alvarez
- Device status
- Awaiting repair
- Location
- 4 West, bay 12
Symptoms
Alarm within 2 minutes at 25 mL/h; line patent, no kinks.
alarm-screen.jpg
02Update ticket
03Audit history
5 entries03 Nov 11:15 · Rafael Alvarez
StatusUnder investigationAwaiting parts
Sensor assembly 81-2210 ordered, ETA 7 Nov. Loaner BME-04230 issued.
03 Nov 10:52 · Rafael Alvarez
Investigation note
Downstream pressure sensor reads 40% high in service mode. Matches KB-112.
03 Nov 10:30 · Rafael Alvarez
StatusAssignedUnder investigation
Alarm reproduces on the test set.
03 Nov 08:05 · Dana Whitfield
StatusReportedAssigned
Assigned to Rafael Alvarez. Two spares left on the unit.
03 Nov 07:42 · 4 West charge nurse
StatusReported
Reported from 4 West, bay 12.