All tickets

DWG 04 · Work order WO-1184

WO-1184REM-02

Under investigationhigh severityon BME-06003 · Electrosurgical unit
  1. Reported
  2. Under investigation
  3. Assigned
  4. In repair
  5. Awaiting parts
  6. Resolved
  7. Closed

01Report

Intermittent REM alarm mid-case, unit swapped for OR 1 spare.

Reported
29 Oct 13:12
Reporter
OR 2 circulating nurse
Error code
REM-02
Assigned to
Mei Chen
Device status
Awaiting repair
Location
OR 2

Symptoms

Return electrode alarm despite good pad contact.

02Update ticket

03Audit history

3 entries
  1. 30 Oct 09:15 · Mei Chen

    StatusAssignedUnder investigation

    Reproduced with a known-good pad. Checking REM board.

  2. 29 Oct 14:00 · Dana Whitfield

    StatusReportedAssigned

    Assigned to Mei Chen.

  3. 29 Oct 13:12 · OR 2 circulating nurse

    StatusReported

    Reported from OR 2.