Registry
DWG 02 · Device recordOperational

Bedside patient monitor

Corvent PM-12 · Intensive Care Unit · ICU, bed 1

12-lead ECG, SpO2, NIBP, 2x IBP. Central station linked.

Biomedical engineering · Asset

BME-05101

S/N PM12-33901

NEXT PM 09 OCT 2027due in 337 days
Category
Patient monitor
Manufacturer
Corvent
Model
PM-12
Serial
PM12-33901
Department
ICU · Intensive Care Unit
Technician
Mei Chen
Purchased
20 Jan 2022
Installed
14 Feb 2022
Warranty
14 Feb 2027
Location
ICU, bed 1
Manual
No document attached

LOGMaintenance history

  1. 1
    09 Oct 2026preventivepassMR-2288

    Annual PM: NIBP calibration, SpO2 simulator, ECG leads, alarms.

    Findings:
    NIBP within 2 mmHg. Hose cracked at connector.
    Parts:
    NIBP hose 71-0021

    Mei ChenNext due 09 Oct 2027Logged 09 Oct 11:00

TICKETSMalfunction reports

0

No malfunctions reported for this device.

REFRelated knowledge base