Registry
DWG 02 · Device recordOperational

C-arm fluoroscopy

Brightline CA-9 · Surgical Services · OR 4

Orthopaedic and vascular cases. Dose reports to PACS.

Biomedical engineering · Asset

BME-01022

S/N CA9-04431

NEXT PM 27 OCT 2027due in 355 days
Category
X-ray
Manufacturer
Brightline
Model
CA-9
Serial
CA9-04431
Department
OR · Surgical Services
Technician
Rafael Alvarez
Purchased
17 Jun 2021
Installed
12 Jul 2021
Warranty
Expired 12 Jul 2026
Location
OR 4
Manual
No document attached

LOGMaintenance history

  1. 1
    27 Oct 2026inspectionpassMR-2293

    Annual radiation safety survey and collimator check.

    Findings:
    Collimator light field within 1.5% of SID.
    Parts:
    None

    Rafael AlvarezNext due 27 Oct 2027CA9-RS-survey.pdfLogged 27 Oct 14:45

TICKETSMalfunction reports

0

No malfunctions reported for this device.

REFRelated knowledge base

No articles for this model yet.

Reference material. Verify with qualified personnel before use.